| Problem Reports |
|---|
| PR Number | Times | Time Lost | Description | Response |
|---|
| 4390 | 02:45-03:00 | 00:15 | ACQ and tip/tilt cameras not responding to commands. Commands turning green in the command status box. | Called: Randy
Shut down pas server, Power off all in PFIP Control tab, Power on all, Start Pas server back up. This worked. |
| 4135 | 23:32-00:17 | 00:45 | During pre night check list got an HFC fault with a prof light on LX track motor. Message in tracker status tab reads MSG LX Track Motor Current Fault. | |
| 4369 | 05:11-05:12 | 00:01 | Pas server slowed to a crawl. 1.0 second exposures taking 30 seconds to update. Top did not show abnormal cpu usage. | Shut down pas server and restarted it. |
| 4391 | 05:13-05:30 | 00:17 | Tracking on sky not fluid. Star jumping around alot with no offset commanded. Date +%s 1455168105. See PR for much more detail. | |
| 4389 | 07:46-07:46 | 00:00 | 6 sensors for segment 54 left out for tonight. They have been removed for 384.00 errors. | |