| Problem Reports |
|---|
| PR Number | Times | Time Lost | Description | Response |
|---|
| 1696 | 00:00-00:00 | 00:00 | removed sensors for LTSE | |
| 4354 | 00:01-00:10 | 00:09 | TCS gui stating CWFS camera offline; no connection to pipeline or images coming through | Did not see any issues or errors had to kill -6 all servers a few times before camera started taking images. |
| 4246 | 00:14-00:20 | 00:06 | TCS gui stating PV camera offline and not responding; | had to restart servers a few times with kill -6 before camera started taking images. |
| 4404 | 00:20-01:11 | 00:51 | TCS server not communicating with RA side or with any syscmd -T commands on htcs; TCS gui stating communication was good to TCS server | Called: Randy , Jason
Called Randy who took a look at it and then we restarted all the servers and started tcs by itself; resolved the issue; issue still unclear. |
| 4405 | 01:40-03:06 | 01:26 | Unable to find M29 while stacking; optics team worked on it during the day | Called: Katie
Katie came up and aligned her first mirror from CCASS tower; mirror was out about -36 tip and 25 tilt. far out of the range of HEFI. |
| 4266 | 03:49-05:19, 05:36-06:11 | 02:05 | ACQcam unable to take images; acqcam is not responding to pipeline or image commands on tcs; | Called: Jim, Randy, Matthew
Seems that acqcam is unable to take images at current temp set point; have to warm up camera for over an hour in order for camera to respond but it fails shortly there after... takes about 30-60 images before it becomes non-responsive. Possible hardware issue. |
| 4406 | 02:00-02:00 | 00:00 | can not execute tcs and other systems command from htcs | |